Error ID: 7890
The automatic wavelength check after lamp ignition has failed. When the lamp is switched on, the detector waits 1 min to warm-up the lamp. Then a check of the deuterium emission line (656 nm) via the reference diode is performed. If the emission line is more than 3 nm away from 656 nm, the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | Calibration incorrect. | Recalibrate the detector. |