Error ID: 7451
The lamp anode voltage is missing. The processor continually monitors the anode voltage across the lamp during operation. If the anode voltage falls below the lower limit, the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | Defective or non-Agilent lamp. | Exchange the lamp. |
2 | Defective UV lamp or non-Agilent lamp. | Exchange the UV lamp. |
3 | Defective power supply. | Please contact your Agilent service representative. |
4 | Defective mainboard. | Please contact your Agilent service representative. |