Error ID: 205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the foam is removed, the fan is switched off, and the error message is generated.
The sensor on the main board detects when the top foam is in place. If the foam is removed during operation, the lamp and grating drive power is switched off, and the error message is generated.
The sensor on the main board detects when the top foam is in place. If the foam is removed, the pump is turned OFF within 4s for safety reasons, and the error message is generated. If the pump is started without cover, this error message is also generated and the pump is not turned on.
Probable cause | Suggested actions | |
---|---|---|
1 | The top foam was removed during operation. | Please contact your Agilent service representative. |
2 | The top foam was removed during operation or pump started without cover. | |
3 | Foam not activating the sensor. | Please contact your Agilent service representative. |
4 | Defective sensor or main board. | Please contact your Agilent service representative. |
5 | Rear of the module is exposed to strong direct sunlight. | Ensure that the rear of module is not directly exposed to strong sunlight. |