For details about the needed parts and orientation please refer to Valve Drive Parts.
For details about the needed parts and orientation please refer to --- MISSING LINK ---.
For details about the needed parts and orientation please refer to --- MISSING LINK ---.
Disassembling and reassembling the valve head
| For bio-inert modules use bio-inert parts only! Do not mix with bio / biocompatible parts. |
| For biocompatible modules use bio / biocompatible parts only! Do not mix with bio-inert parts. |
When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| Hex key |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| Quick Change Valve Head |
WARNING
Toxic, flammable and hazardous solvents, samples and reagents
The handling of solvents, samples and reagents can hold health and safety risks.
Use the Hex Key to open and remove the Stator Screws from the Stator Head.
Carefully disassemble the necessary Valve Head parts to gain access to the one you want to replace. While doing so please observe the orientation of the parts.
Independent of the part you want to replace always inspect all parts for signs of damage.
Replace the proposed part.
NOTE
Always mind the correct orientation of the parts and avoid to touch their surfaces.
Turn each of the screws an equal amount until they are finger-tight, then tighten them for another half turn.
base-id: 3182152587
id: 9007202436893579