Error ID: 4526
Probable cause | Suggested actions | |
---|---|---|
1 | Communication issues. | Check the CAN cable connections of the modules. |
2 | Configuration mismatch. | Check and correct if necessary the valve configuration and presence of defined position cluster partner. If the module was moved to another LC stack, perform Firmware Declustering in Service & Diagnostic section of Lab Advisor. |
base-id: 3182068107