Error ID: 2523
The connection from the valve drive to a defined pressure cluster partner is lost.
Probable cause | Suggested actions | |
---|---|---|
1 | Communication issues. | Check the CAN cable connections of the modules. |
2 | Configuration mismatch. | Check and correct if necessary the valve configuration and presence of defined pressure cluster partner. |
base-id: 3182046731