Error ID: 24006
The valve drive identified a different valve head than it had identified during the last initialization.
NOTE
Soft power-down power supply of the valve drive.
Whenever you want to power cycle the valve drive for a re-boot, it needs to be powered off for at least 10 seconds.
Probable cause | Suggested actions | |
---|---|---|
1 | A valve head has been exchanged (hot-plugged) while the valve drive was still powered on. | Change the valve head. It is important to have the valve switched off for at least 10 s after or before a new valve head has been installed. |
base-id: 3182033803
id: 9007202436774795