Error ID: 24001
The valve drive was not able to operate the valve head correctly. Either due to mechanical reasons or the movement couldn’t be detected correctly.
Probable cause | Suggested actions | |
---|---|---|
1 | Mechanical problems. Friction too high or blockages on the valve drive’s motor or on the valve head. | Check valve head for correct installation. Try to identify the source of trouble by installing a different valve head if possible. Please contact your Agilent service representative. |
2 | Defect Sensor on the Valve Drive Motor. | Check valve head for correct installation. Try to identify the source of trouble by installing a different valve head if possible. Please contact your Agilent service representative. |
base-id: 3181995531