The Agilent InfinityLab Quick Change Valves can be installed in the Multicolumn Thermostat (G7116A/B), the 1290 Infinity Flexible Cube (G4227A) and in the 1290 Infinity III Valve Drive (G1170A).
| For bio-inert modules use bio-inert parts only! Do not mix with bio / biocompatible parts. |
| For biocompatible modules use bio / biocompatible parts only! Do not mix with bio-inert parts. |
Valve Kit PN | Pressure [bar] | Positions | Ports | Typical Application |
---|---|---|---|---|
G4231A | 800 | 2 | 6 | Any two-way switching, e.g. between two detectors, between waste and detector, between two columns |
G4231C | 1300 | 2 | 6 | |
G4232C | 800 | 2 | 10 | Anything a 2-position/6-port valve can do plus alternating column regeneration. |
G4232D | 1300 | 2 | 10 | |
G4234A | 800 | 6 | 14 | 6 column selection |
G4234C | 1300 | 6 | 14 | |
G4235A | 210 | 12 | 13 | Solvent selection or fractionation |
G4237A | 800 | 4 | 10 | 4 column selection |
G4239C | 1300 | 8 | 18 | 8 column selection |
G5631A | 600 | 2 | 6 | Bio-inert, same as G4231A |
G5632A | 600 | 2 | 10 | Bio-inert, anything a 2-position/6-port valve can do plus alternation column regeneration |
G5639A | 600 | 4 | 10 | Bio-inert 4 column selection |
G5641A | 1300 | 2 | 10 | Bio, anything a 2-position/6-port valve can do plus alternation column regeneration |