Error ID: 32031
Switching of the valve failed.
Probable cause | Suggested actions | |
---|---|---|
1 | Mechanical problems. Friction too high or blockages on the valve drive’s motor or on the valve head. | Check valve head for correct installation. Try to identify the source of trouble (valve head or drive) by installing a different valve head if possible. Please contact your Agilent service representative. |
2 | Defective sensor on the valve drive motor | Check valve head for correct installation. Try to identify the source of trouble (valve head or drive) by installing a different valve head if possible. Please contact your Agilent service representative. |
base-id: 3002720395