Error ID: 32030
During initialization, the valve actuator turns until the encoder reads the reference index position. An error is generated, if the reference index cannot be found within a given time.
Probable cause | Suggested actions | |
---|---|---|
1 | Defect in cable connection of valve drive control. | Please contact your Agilent service representative. |
2 | Defect in cable connection of valve actuator encoder reader. | Please contact your Agilent service representative. |
3 | Defect of valve drive or valve actuator encoder reader. | Please contact your Agilent service representative. |
base-id: 3002673675
id: 9007202257414667