Error ID: 70
A not-ready condition is still present on the remote input. When an analysis is started, the system expects all not-ready conditions (for example, a not-ready condition during detector balance) to switch to run conditions within one minute of starting the analysis. If a not-ready condition is still present on the remote line after one minute the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | Not-ready condition in one of the instruments connected to the remote line. | Ensure the instrument showing the not-ready condition is installed correctly, and is set up correctly for analysis. |
2 | Defective remote cable. | Exchange the remote cable. |
3 | Defective components in the instrument showing the not-ready condition. | Check the instrument for defects (refer to the instrument’s documentation). |