Error ID: 71
During an analysis, the internal synchronization or communication between one or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one or more of the modules is no longer recognized as being connected to the system, the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | CAN cable disconnected. | Ensure all the CAN cables are connected correctly. Ensure all CAN cables are installed correctly. |
2 | Defective CAN cable. | Exchange the CAN cable. |
3 | Defective mainboard in another module. | Switch off the system. Restart the system, and determine which module or modules are not recognized by the system. |
base-id: 3002380683
id: 9007202257121675