Error ID: 62
The timeout threshold was exceeded.
Probable cause | Suggested actions | |
---|---|---|
1 | The analysis was completed successfully, and the timeout function switched off the module as requested. | Check the logbook for the occurrence and source of a not-ready condition. Restart the analysis where required. |
2 | A not-ready condition was present during a sequence or multiple-injection run for a period longer than the timeout threshold. | Check the logbook for the occurrence and source of a not-ready condition. Restart the analysis where required. |
base-id: 3002359307