| For biocompatible modules use bio / biocompatible parts only! Do not mix with bio-inert parts. |
| For bio-inert modules use bio-inert parts only! Do not mix with bio / biocompatible parts. |
Flow rate | 0.075 mm i.d. capillary | 0.12 mm i.d. capillary | 0.17 mm i.d. capillary |
---|---|---|---|
< 2 mL/min | Ultra‐low Dispersion G7116‐60021 (Internal volume: 1.0 µL) | Standard Flow G7116‐60015 (Internal volume: 1.6 µL) | Large ID G7116‐60051 (Internal volume: 3.0 µL) |
> 2 mL/min |
| High Flow G7116‐60031 (Internal volume: 3.0 µL) | Large ID High Flow G7116‐60061 (Internal volume: 6.0 µL) |
Bio, all flow rates | Bio Ultra-low Dispersion G7116-60091 (Internal volume: 1.0 µL) | Bio Standard Flow G7116-60071 (Internal volume: 1.6 µL) Bio High Flow G7116-60081 (Internal volume: 3.0 µL) |
|
Bio-inert, all flow rates |
|
| Bio‐inert G7116‐60041 (Internal volume: 9.0 µL) |
Quick Connect Heat Exchangers with 0.12 mm and 0.075 mm i.d. capillary are suitable for G7116B. Quick Connect Heat Exchangers with 0.17 mm i.d. are suitable for G7116A.
Item | Description |
---|---|
| Old design Quick Connect Heat Exchanger: twist lock clips of the Quick Connect Heat Exchanger are used to secure the Quick Connect Heat Exchanger in the heater block of the MCT. |
| New design Quick Connect Heat Exchanger: column holders (lamella type G7116-68003 or clamp type G7116-68004) are used to secure the Quick Connect Heat Exchanger in the heater block of the MCT. G7116-68003 lamella type column holders are delivered as a part of the Quick Connect Heat Exchanger and do not need to be ordered separately. |
| Design has been changed for all Quick Connect Heat Exchangers except G7116-60041, Quick Connect Heat Exchanger Bio-inert. |