| For bio-inert modules use bio-inert parts only! Do not mix with bio / biocompatible parts. |
| For biocompatible modules use bio / biocompatible parts only! Do not mix with bio-inert parts. |
Modified design of Quick Connect Heat Exchangers allows simplified installation of the heat exchanger in the Multicolumn Thermostat (MCT).
New design Quick Connect Heat Exchangers do not have twist lock clips on the front side of the heat exchanger. Instead, Quick Connect Heat Exchangers include two column holders to fix the heat exchanger in the MCT heater block.
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 5043-0915 | Fitting mounting tool |
1 | ![]() | 5023-2502 | Hex driver SW-6.35, slitted |
2 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
Parts required
For 1290
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G7116-60015 | Quick Connect Heat Exchanger Standard |
1 | ![]() | G7116-60021 | Quick Connect Heat Exchanger Ultra Low Dispersion |
1 | ![]() | G7116-60031 | Quick Connect Heat Exchanger High Flow |
1 | ![]() | 5067-5957 | InfinityLab Quick Connect Assy ST 0.12 mm x 105 mm (from heat exchanger to column) |
1 | ![]() | G7116-68003 | Column Holder Lamella, 2/pk |
Parts required
For 1260
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G7116-60061 | Quick Connect Heat Exchanger Large ID High Flow |
1 | ![]() | 5500-1193 | InfinityLab Quick Turn Capillary ST 0.17 mm x 105 mm, long socket (from heat exchanger to column) |
1 | ![]() | G7116-60051 | Quick Connect Heat Exchanger Large ID |
1 | ![]() | 5067-5966 | InfinityLab Quick Turn Fitting |
1 | ![]() | G7167-68703 | Fitting Intermediate Kit |
1 | ![]() | G7116-68003 | Column Holder Lamella, 2/pk |
Parts required
Bio-inert option
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G7116-60041 | Quick Connect Heat Exchanger Bio-inert |
1 | ![]() | 5067-4780 | Bio Capillary 0.17 mm x 300 mm, 1.6 mm OD socket (from column to detector) |
1 | ![]() | 0100-1516 | Finger-tight fitting PEEK, 2/pk |
1 | ![]() | 5067-4741 | ZDV union (Bio-inert) |
1 | ![]() | 5067-5403 | UHP fitting |
1 | ![]() | G7116-68003 | Column Holder Lamella, 2/pk |
Parts required
Bio-compatible option
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G7116-60071 | Quick Connect Bio Heat Exchanger Standard Flow |
1 | ![]() | G7116-60081 | Quick Connect Bio Heat Exchanger High Flow |
1 | ![]() | G7116-68003 | Column Holder Lamella, 2/pk |
1 | ![]() | 5500-1596 | Quick Turn Capillary MP35N 0.12 mm x 280 mm |
1 | ![]() | 5067-5966 | InfinityLab Quick Turn Fitting |
1 | ![]() | 5500-1578 | Quick Connect Capillary MP35N 0.12 mm x 105 mm |
1 | ![]() | 5067-5965 | InfinityLab Quick Connect LC fitting |
Possible positions for heat exchangers. If only one column is used, the preferred positions are 2, 3, 6, or 7.
Connect the sampler outlet capillary (or the capillary from the column selection valve) to the inlet port of the heat exchanger.
Connect the column connection capillary to the outlet of the Quick Connect Heat Exchanger.
Position the heat exchanger in the groove of the MCT heater block and prepare column holders to fix the heat exchanger.
Turn column holders to fix the Quick Connect Heat Exchanger in the MCT.
NOTE
The column holder clip can be mounted at any free spot on top of the heat exchanger.