Type | Specification | Comments |
---|---|---|
Weight | 11.0 kg (24.3 lbs non-cooled) |
|
Dimensions (height × width × depth) | 420 x 200 x 450 mm (16.5 x 7.9 x 17.7 inches) |
|
Line voltage | 100 – 120 / 220 – 240 V~, ± 10 % |
|
Line frequency | 50 or 60 Hz, ± 5 % |
|
Power consumption | 150 W, 2 A (max) |
|
Ambient operating temperature | 10 – 35 °C (50 – 95 °F) |
|
Ambient non-operating temperature | -40 – 70 °C (-40 – 158 °F) |
|
Humidity | < 10 – 80 % r.h at 40° C (104° F) | Non-condensing |
Operating altitude | Up to 2000 m (6562 ft) |
|
Safety standards: IEC, EN, CSA, UL | Overvoltage category II, Pollution degree 2 | For indoor use only |
ISM classification | ISM Group 1 Class B | According to CISPR 11 |
Type | Specification | Comment |
---|---|---|
Light Source | LED 480 nm (Class 1 LED product) |
|
Detector | PMT with digital signal processing |
|
Nebulizer | OFF, 25 – 90 °C |
|
Designed for use with Agilent InfinityLab Assist | Intuitive User Interface, Automated Workflows, Predictive Maintenance & Assisted Troubleshooting |
|
Evaporator |
|
|
Non-cooled | OFF, 25 – 120 °C |
|
Cooled | Not Applicable |
|
Gas Flow Range | 0.9 – 3.25 SLM (controlled gas shut-off) |
|
Short Term Noise | < 0.2 mV | See conditions in Specification Conditions of the 1260 Infinity III Evaporative Light Scattering Detector (G4260B). |
Drift | < 1 mV/h | See conditions in Specification Conditions of the 1260 Infinity III Evaporative Light Scattering Detector (G4260B). |
Operating Pressure | 4.1 – 6.9 bar (60 – 100 psi) |
|
Eluent Flow range | 0.2 – 5.0 mL/min |
|
Digital Output | 10, 40 or 80 Hz ( 24 bit) |
|
Analogue Output | 0 – 1.25 V FSD |
|
Communication | Ethernet Serial (RS232) Remote Start Input Pump Stop: 1 Contact closure |
|
Remote operation | Remote Start Input |
|
Instrument control | ELSD driver for OpenLab CDS ELSD driver for OpenLAB ChemStation edition ELSD driver for OpenLAB EZChrom edition LC and CE Drivers Rev. A.02.11 or above for ICF | InfinityLab Assist (G7180A) with firmware D.07.40 or above not supported! For details about supported software versions refer to the compatibility matrix of the ELSD compatibility matrix and ICOCO |
Safety and maintenance | Gas shut-off Valve, Leak Detection | No Laser fitted |
NOTE
In order to control OpenLab CDS, the ELSD driver is required.