Error ID: 81
The ambient-compensation sensor (NTC) on the mainboard in the module has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the mainboard is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes. If the resistance across the sensor increases above the upper limit, the error message is generated.
The ambient-compensation sensor (NTC) on the on/off switch board in the module has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the on/off switch board is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes. If the resistance across the sensor increases above the upper limit, the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | Loose connection between the on/off switch board and the mainboard. | Please contact your Agilent service representative. |
2 | Defective on/off switch assembly. | Please contact your Agilent service representative. |
3 | Defective mainboard. | Please contact your Agilent service representative. |
base-id: 3002622475
id: 63050397785809419