Error ID: 82
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak sensor current to change within defined limits. If the current increases above the upper limit, the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | Defective leak sensor. | Please contact your Agilent service representative. |
2 | Leak sensor incorrectly routed, being pinched by a metal component. | Please contact your Agilent service representative. |
3 | On/Off switch assembly defective. | Please contact your Agilent service representative. |
4 | Cable or contact problem. | Please contact your Agilent service representative. |
base-id: 3002453003