When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G4267-87201 | Needle Assembly OR |
1 | ![]() | G4267-87210 | Needle Assembly (slotted) for high injection volumes |
Prerequisites
|
WARNING
Risk of injury by uncovered needle
An uncovered needle is a risk of harm to the operator.
NOTE
It is recommended to always exchange the needle assembly and the needle seat at the same time to prevent premature leakage.
CAUTION
Blockages inside of the needle assembly union
Use a 1/4 inch wrench to tighten the fitting of the loop capillary.
In the Local Controller close Change needle /seat.
OR
In the Agilent Lab Advisor software Change needle/loop , click Back and wait until the needle assembly is in the needle park station.
Perform a Pressure Test.