The Agilent 1260 Infinity II SFC System consists of the modules shown below.
The 1260 Infinity II SFC Control Module, redistills and preconditions the CO₂ by boosting he pressure to just below the column head pressure, relieving the 1260 Infinity II SFC Binary Pump of any compression requirements. This results in low detector noise and significantly higher sensitivity. Therefore, the 1260 Infinity II SFC Binary Pump receives pre-conditioned CO₂, and acts only as metering device for the mobile phase flow and to form the gradient with the second pump head by adding the appropriate amount of modifier solvent. Downstream of the detector, the mobile phase is redirected back into the 1260 Infinity II SFC Control Module to an integrated back-pressure regulator that maintains the back-pressure over the system. The Agilent 1260 Infinity II SFC System is completely controlled by Agilent OpenLab CDS software.