When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]()
|
| Hex key 4 mm (part of 5023-2524Hex Key Set) |
1 | ![]()
|
| Cleaning tissue and appropriate solvent like isopropanol or methanol |
Prerequisites
|
In the Local Controller start the maintenance mode and select Change Metering Device function.
OR
In the Agilent Lab Advisor software select Service & Diagnostics in the system screen Maintenance Positions > Change Metering Device , click Start and wait until the metering device is in maintenance position.
Carefully remove the seal from the tip of the piston.