Safety Information Related to Maintenance
WARNING
Eye discomfort
The light source in the G4260B ELSD is a Class 1 LED product.
Temporary discomfort may result from directly viewing the light produced by
this source.
- Do not look into the
beam.
WARNING
Eye damage by detector light
Eye damage may result from directly viewing the UV-light produced by the lamp of the optical system used in this product.
- Always turn the lamp of the optical system off before removing it.
WARNING
Fire
and damage to the module
Wrong
fuses
- Make sure that only fuses with the
required rated current and of the specified type (super-fast, fast, time delay
etc) are used for replacement.
- The use of repaired fuses and the
short-circuiting of fuse-holders must be
avoided.
WARNING
Heavy weight
The module is heavy.
- Carry the module at least with 2 people.
- Avoid back strain or injury by following all precautions for lifting heavy objects.
- Ensure that the load is as close to your body as possible.
- Ensure that you can cope with the weight of your load.
WARNING
Personal injury or damage to the product
Agilent is not responsible for any damages caused, in whole or in part,
by improper use of the products, unauthorized alterations, adjustments or
modifications to the products, failure to comply with procedures in Agilent
product user guides, or use of the products in violation of applicable laws,
rules or regulations.
- Use your Agilent
products only in the manner described in the Agilent product user
guides.
WARNING
Electrical shock
Repair work at
the module can lead to personal injuries, e.g. shock hazard, when the cover is
opened.
- Do not remove the cover of the
module.
- Only certified persons are authorized to
carry out repairs inside the module.
WARNING
Sharp metal edges
Sharp-edged parts of the equipment may cause injuries.
- To prevent personal injury, be careful when getting in contact with sharp metal areas.
WARNING
Hot heat exchangers

The column compartment has two heat exchanger assemblies that might be hot.
- Allow them to cool down before starting repairs.
WARNING
Toxic, flammable and hazardous solvents,
samples and reagents
The handling of solvents,
samples and reagents can hold health and safety risks.
- When working with these substances observe appropriate safety
procedures (for example by wearing goggles, safety gloves and protective
clothing) as described in the material handling and safety data sheet supplied
by the vendor, and follow good laboratory practice.
- The volume of substances should be reduced to the minimum required for
the analysis.
- Do not operate the instrument in an
explosive atmosphere.
CAUTION
Safety standards for external equipment
- If
you connect external equipment to the instrument, make sure that you only use
accessory units tested and approved according to the safety standards
appropriate for the type of external
equipment.
CAUTION
Sample degradation and contamination of the instrument
Metal parts in the flow path can interact with the bio-molecules in the sample leading to sample degradation and contamination.
- For bio applications, always use dedicated bio parts, which can be identified by the bio-inert symbol or other markers described in this manual.
- Do not mix bio, and non-bio modules or parts in a bio system.
base-id: 3873298699
id: 54043199401744651-3