Error ID: 83
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak sensor current to change within defined limits. If the current falls outside the lower limit, the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | Leak sensor not connected to the mainboard. | Please contact your Agilent service representative. |
2 | Leak sensor not connected to the on/off switch board. | Please contact your Agilent service representative. |
3 | Defective leak sensor. | Please contact your Agilent service representative. |
4 | Leak sensor incorrectly routed, being pinched by a metal component. | Please contact your Agilent service representative. |
5 | On/Off switch assembly defective. | Please contact your Agilent service representative. |