When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
Prerequisites
|
WARNING
Risk of injury by uncovered needle
An uncovered needle is a risk of harm to the operator.
In the Agilent Lab Advisor software select Service & Diagnostics > Maintenance Positions > Change Needle, Loop and Seat , click Start and wait until the needle assembly is in maintenance position.
OR: In the Local Controller start the maintenance mode and select Change Needle, Loop and Seat function.
The needle assembly is still connected to the loop capillary. Use a 1/4 inch wrench to loosen the fitting of the loop capillary connected to the analytical head.
Lock the needle in the safety position.
CAUTION
Damage of the loop
The loop shape may be damaged if the loop is stretched or bent too far.
WARNING
Sharp needle
Uncovered needles may cause injuries
Remove the needle assembly by slightly pulling the needle cartridge.
Remove the cartridge out of its proper position. By gently tilting and pulling it out of the work space of the multisampler.
base-id: 4548160139
id: 4548160139