The SFC Control Module has a 1/8 inch tube inlet connection on the lower left side on the back of the module. This connection is actually part of a very high surface area filter intended to intercept catastrophic levels of particulates. You need to connect this input to a source of carbon dioxide. The most common source of carbon dioxide is liquefied carbon dioxide from a room-temperature cylinder. At room temperature, the pressure in the cylinder could change from a little above 50 bar to just below 70 bar. Unlike most SFCs, Agilent's G4301A SFC Control Module is immune to the variations in flow resulting from cylinder pressure.
WARNING
Creating severe frostbite in a short time
WARNING
Use the system in a well ventilated area
base-id: 4540245643
id: 4540245643