The temperature of the mobile phase prior to detection is a critical parameter for minimizing baseline noise recorded in the detector flow cell. The heat exchanger on the right side of the column compartment is used to pre- heat the mobile phase before it enters the column, indirectly heating the column. The heat exchanger on the left side is used to change the temperature of the mobile phase to achieve minimum noise. This is of crucial importance, since the refractive index of carbon dioxide responds up to 50 times stronger than water- based mobile phases; thus, even small changes in temperature can significantly affect noise levels. Temperature changes of mobile phase with the left side heat exchanger e.g. between 38 °C and 49 °C resulted in a variation in peak-to-peak noise of over an order of magnitude.
The recommended temperature setting on the left heat exchanger in order to achieve minimum noise is typically between 37 °C and 40 °C for G7115A or G7165A detectors.