When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-1924 | Open-end wrench 14 mm |
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]() | 5067-5688 | Torque wrench 1 - 25 Nm with 14 mm wrench |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G1312-60067 | Outlet valve , OR |
1 | ![]() | G1312-60167 | Outlet Valve Type N/SFC |
Prerequisites
|
Using a ¼ inch wrench disconnect the absorber capillary from the outlet valve.
Unscrew the valve with the 14 mm wrench and remove it from the pump body.
NOTE
Do not disassemble the outlet valve, as this can damage the valve.
Reinstall the outlet valve and tighten it using a torque wrench (approx.12 Nm).
Reconnect the capillary.