When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]()
|
| Hexagonal key, 4 mm |
1 | ![]() | 5023-0240 | Hex driver, ¼", slitted |
Prerequisites
|
CAUTION
Damage of the pump drive
Starting the pump when the pump head is removed may damage the pump drive.
NOTE
Both pump head assemblies use the same internal components. In addition, pump head A is fitted with the purge valve. The following procedure describes the removal and disassembly of pump head A (left). For pump head B (right) proceed in the same way and skip steps that deal with the purge valve.
Open the doors.
Disconnect the capillaries. Beware of leaking solvents.
Using a 4 mm hexagonal key stepwise loosen and remove the two pump head screws.
Remove the pump head assembly from the module.