Type | Specification | Comment |
---|---|---|
Weight | 8.9 kg (19.6 lbs) |
|
Dimensions (height × width × depth) | 169 × 396 × 441 mm (6.7 × 15.6 × 17.4 inches) |
|
Line voltage | 100 – 240 V~ | Wide-ranging capability |
Line frequency | 50 / 60 Hz |
|
Power consumption | 20 VA / 20 W |
|
Ambient operating temperature | 4–40 °C (39–104 °F) |
|
Ambient non-operating temperature | -20–70 °C (-4–158 °F) |
|
Humidity | < 95% r.h. at 40 °C (104 °F) | Non-condensing |
Operating altitude | Up to 3000 m (9842 ft) |
|
Safety standards: IEC, EN, CSA, UL | Overvoltage category II, Pollution degree 2 | For indoor use only. |
ISM Classification | ISM Group 1 Class B | According to CISPR 11 |
Type | Specification | Comment |
---|---|---|
Designed for use with Agilent InfinityLab Assist | Intuitive User Interface, Automated Workflows, Predictive Maintenance & Assisted Troubleshooting |
|
Supported solvent bottles | 0.125 - 2.5 L with a maximum bottle diameter of 100 mm (small positions) and 150 mm (large positions) |
|
Number of bottles | 4 solvent bottles up to 100 mm diameter (1 L bottles), and 2 up to 150 mm diameter (2.5 L bottles); 2 bottles of max diameter of 100 mm can be placed in addition between the weighing positions without solvent sensing |
|
Maximum load weight per position | 5 kg |
|
Minimum load weight per position | 50 g |
|
Instrument control | LC & CE Drivers 3.9 or above InfinityLab Assist (G7180A) with firmware D.07.42 or above Lab Advisor B.02.22 or above | For details about supported software versions refer to the compatibility matrix of your version of the LC & CE Drivers |
Communication | Controller Area Network (2 x CAN Port, RJ45) | G7175A is a hosted module. (The LC stack needs to contain a suitable host module for communication and control) |
Maintenance and safety-related features | Extensive diagnostics, error detection and display with Agilent InfinityLab Assist and Agilent Lab Advisor. Leak detection, safe leak handling, leak output signal for shutdown of pumping system, and low voltages in major maintenance areas |
|
GLP features | Early maintenance feedback (EMF) for continuous tracking of instrument usage with user-settable limits and feedback messages. Electronic records of maintenance and errors |
|
Housing | All materials recyclable |
|
base-id: 10443089931