Type | Specification | Comments |
---|---|---|
Weight | 11.5 kg (25.4 lbs) |
|
Dimensions (height × width × depth) | 140 x 396 x 436 mm (5.5 x 15.6 x 17.2 inches) |
|
Line voltage | 100–240 V~, ±10% | Wide-ranging capability |
Line frequency | 50 or 60 Hz, ±5% |
|
Power consumption | 110 VA, 100 W |
|
Ambient operating temperature | 4–55 °C (39–131 °F) |
|
Ambient non-operating temperature | -40–70 °C (-40–158 °F) |
|
Humidity | < 95% r.h. at 40 °C (104 °F) | Non-condensing |
Operating altitude | Up to 3000 m (9842 ft) |
|
Safety standards: IEC, EN, CSA, UL | Overvoltage category II, Pollution degree 2 | For indoor use only |
ISM classification | ISM Group 1 Class B | According to CISPR 11 |
Type | Specification |
---|---|
Detection type | 1024-element photodiode array |
Designed for use with Agilent InfinityLab Assist | Intuitive User Interface, |
Light source | Deuterium lamp |
Number of signals | 8 |
Maximum data rate | 240 Hz (both spectra and signals) |
Short term signal noise (ASTM) | With 10 mm Max-Light cartridge cell: <±3·10-6 AU/cm at 230 nm, slit width 4 nm, TC 2 s With 60 mm Max-Light cartridge cell: <±0.6·10-6 AU/cm at 230 nm, slit width 4 nm, TC 2 s |
Drift | < 0.5·10-3 AU/h at 230 nm |
Linear absorbance range | > 2 AU (5 %) at 273 nm, typically 2.5 AU (5 %) |
Wavelength range | 190 – 640 nm |
Wavelength accuracy | ± 1 nm, self-calibration with deuterium lines |
Wavelength precision | < ± 0.1 nm |
Wavelength bunching | 2 – 400 nm, programmable in 1 nm steps |
Slit width | 1, 2, 4, 8 nm |
Diode width | ~0.5 nm |
Nominal spectral resolution | < 1 nm |
Time programmable | Wavelength, polarity, peak width, lamp bandwidth, auto balance, wavelength range, threshold, spectra storage mode |
Flow cells | User-exchangeable, self-aligning cartridge cells with RFID tags. Max-Light Cartridge Cell (Standard): 10 mm, σV = 1.0 µL Max-Light Cartridge Cell (High Sensitivity): 60 mm, σV = 4 µL Max-Light Cartridge Ultra Low Dispersion (ULD) Cell: 10 mm, σV = 0.6 µL Max-Light Cartridge High Dynamic Range (HDR) Cell: 3.7 mm, σV = 0.9 µL Max-Light Cartridge Cell LSS: 10 mm, σV = 1.0 µL Maximum Operating Pressure (MOP)Maximum operating pressure (MOP): 70 bar Maximum Incidental Pressure (MIP)Maximum incidental pressure (MIP): 150 bar |
Spectral tools | Data analysis software for spectra evaluation, including spectral libraries and peak purity functions |
Analog output | Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, one output |
Instrument control | LC and CE Drivers A.02.11 or above Instrument Control Framework (ICF) A.02.04 or above Lab Advisor B.02.06 or above InfinityLab Assist (G7180A) with firmware D.07.40 or above Agilent Instant Pilot (G4208A) with firmware B.02.19 or above For details about supported software versions refer to the compatibility matrix of your version of the LC and CE Drivers |
Communication | LAN, Controller Area Network (CAN), USB ERI: ready, start, stop and shut-down signals |
GLP | RFID for electronics records of flow cell and UV lamp conditions (path length, volume, product number, serial number, test passed, usage). Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of lamp burn time with user settable limits and feedback messages. Electronic records of maintenance and errors. Verification of wavelength accuracy with deuterium lines. |
Safety and maintenance | Extensive diagnostics, error detection and display through Agilent InfinityLab Assist, Agilent Instant Pilot and Agilent Lab Advisor software. Leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas. |
Housing | All materials recyclable. |
Others | Second generation of Electronic Temperature Control (ETC) for the complete optical unit. |
[1]Maximum operating pressure (MOP): Maximum pressure at which a system can operate continuously under normal conditions.
[2]Maximum incidental pressure (MIP): The maximum pressure which the system can experience during a short time.