Type | Specification | Comments |
---|---|---|
Weight | 22 kg (48.5 lbs) | w/o sample thermostat, w/o sample ID reader |
Dimensions (height × width × depth) | 320 x 396 x 468 mm (12.6 x 15.6 x 18.4 inches) |
|
Line voltage | 100–240 V~, ±10% | Wide-ranging capability |
Line frequency | 50 or 60 Hz, ±5% |
|
Power consumption | 180 VA, 180 W |
|
Ambient operating temperature | 4–40 °C (39–104 °F) |
|
Ambient non-operating temperature | -40–70 °C (-40–158 °F) |
|
Humidity | < 95% r.h. at 40 °C (104 °F)SampleThermostatIncluded | Non-condensing |
Operating altitude | Up to 3000 m (9842 ft) |
|
Safety standards: IEC, EN, CSA, UL | Overvoltage category II, Pollution degree 2 | For indoor use only. |
ISM Classification | ISM Group 1 Class B | According to CISPR 11 |
Permitted solvents | Boiling point ≥ 56 °C (133 °F) Auto-ignition temperature ≥ 200 °C (392 °F) |
|
[1]If a sample thermostat is included the upper value for humidity can be reduced. Please check your lab conditions to stay beyond dew point values for non–condensing operation.
Type | Specification | Comments |
---|---|---|
Designed for use with Agilent InfinityLab Assist | Intuitive User Interface, Automated Workflows, Predictive Maintenance & Assisted Troubleshooting |
|
Injection range | Default: 0.1 – 100 µL in 0.1 µL increments | Up to 600 bar using the 100 µL analytical head |
0.1 – 250 µL or 1000 µL in 0.1 µL increments with G5667-68711 (hardware modification required) | Pressure range up to 275 bar Multidraw mode (injection into needle seat capillary) | |
Injection precision | <0.15 % RSD or SD <10 nL, whatever is greater | Measured caffeine |
Injection linearity | 0.9999 in the range of 0.1 – 100 µL | Measured caffeine |
Pressure range | Up to 600 bar | Max pressure for basic instrument |
Sample viscosity range | 0.2 – 5.0 cP |
|
Sample capacity | 1H Drawer up to 8 drawers and 16 positions shallow well plates (MTP) | Max. 6144/1536 samples (384MTP/96) |
2H Drawer up to 4 drawers and 8 positions MTP, deep well plates, vials, Eppendorf | 3072 samples, 432 vials (2 mL) | |
3H Drawer up to 2 drawers and 4 positions MTP, deep well plates, vials up to 6 mL, Eppendorf | 1536 samples, 60 vials (6 mL), 384 vials (1 mL), 216 vials (2 mL) | |
Injection cycle time | <10 s using following standard conditions: Default draw speed: 100 µL/min Default eject speed: 400 µL/min Injection volume: 1 µL | Time between 2 injections is not mechanically limited, time delay depends on communication speed of software, OS or network connections |
Carryover | <0.003 % (30 ppm) Multisampler Standard <0.0009 % (9 ppm) Multisampler Multiwash | Sample: Chlorhexidine |
Multiwash | Outer needle wash and seat backflush for carryover reduction with up to 3 different solvents |
|
Instrument Control | LC and CE Drivers A.02.14 or above Instrument Control Framework (ICF) A.02.04 or above InfinityLab Assist (G7180A) with firmware D.07.40 or above Lab Advisor B.02.08 or above Instant Pilot (G4208) with firmware B.02.20 or above | For details about supported software versions refer to the compatibility matrix of your version of the LC & CE Drivers |
Communication | Controller Area Network (CAN), Local Area Network (LAN), ERI: ready, start, stop and shut-down signals |
|
Maintenance and safety-related features | Extensive diagnostics, error detection and display with Agilent Lab Advisor software Leak detection, safe leak handling, leak output signal for shutdown of pumping system, and low voltages in major maintenance areas |
|
GLP features | Early maintenance feedback (EMF) for continuous tracking of instrument usage with user-settable limits and feedback messages. Electronic records of maintenance and errors. |
|
Housing | All materials recyclable. |
|
base-id: 5658563339