Type | Specification | Comments |
---|---|---|
Weight | 12.5 kg (27.6 lbs) |
|
Dimensions (height × width × depth) | 160 x 435 x 436 mm (6.3 x 17.1 x 17.2 inches), Width with column identification kit: 472 mm | G7116B may have a Column ID tag reader on both sides. |
Line voltage | 100–240 V~, ±10% | Wide-ranging capability |
Line frequency | 50 or 60 Hz, ±5% |
|
Power consumption | 150 VA, 150 W |
|
Ambient operating temperature | 4–55°C (39–131°F) |
|
Ambient non-operating temperature | -40–70°C (-40–158°F) |
|
Humidity | < 95% r.h. at 40°C (104°F) | Non-condensing |
Operating altitude | Up to 3000 m (9842 ft) |
|
Safety standards: IEC, EN, CSA, UL | Overvoltage category II, Pollution degree 2 | For indoor use only |
ISM Classification | ISM Group 1 Class B | According to CISPR 11 |
Feature | Specification | Comment |
---|---|---|
Operating principle | Thermostatted column compartment with dual, independent Peltier element. Solvent pre-heating and still-air operation for reduction of chromatographic band broadening under UHPLC-conditions. Up to four devices can be clustered and controlled by a single user interface for additional flexibility.Requires LC and CE drivers A.02.12 or above |
|
Designed for use with Agilent InfinityLab Assist | Intuitive User Interface, Automated Workflows, Predictive Maintenance & Assisted Troubleshooting | |
Temperature range | 20 °C below ambient (minimum 4 °C) to 110 °C | All specifications are valid for distilled water at ambient temperature |
Temperature range increment | Settable in steps of 0.1 °C | All specifications are valid for distilled water at ambient temperature |
Temperature stability | ± 0.03 °C | All specifications are valid for distilled water at ambient temperature |
Temperature accuracy | ± 0.5 °C (with calibration at 40 °C) | All specifications are valid for distilled water at ambient temperature |
Temperature precision | 0.05 °C | All specifications are valid for distilled water at ambient temperature |
Independent temperature zones | 2 (in single device) Up to 8 in clustered configurationRequires LC and CE drivers A.02.12 or above |
|
Column capacity | 8 columns of 100 mm length plus Quick Connect fittings or pre-columns 4 columns of 300 mm length plus Quick Connect fittings or pre-columns Selection of columns by single optional integrated 8-column selector valve (1300 bar) Maximum of 32 columns of 100 mm length plus Quick Connect fittings or pre-columns 16 columns of 300 mm length plus Quick Connect fittings or pre-columns with clusteringRequires LC and CE drivers A.02.12 or above of four devices. |
|
Heat-up/cool-down time | 5 min from ambient to 40 °C 10 min from 40 °C to 20 °C <30 min from 25 °C to 100 °C | All specifications are valid for distilled water at ambient temperature |
Solvent heat exchangers | Individually quick-installable for every column. Available at 1 µL internal volume, 0.075 mm i.d. capillary (ultra-low dispersion), 1.6 µL internal volume, 0.12 mm i.d. capillary (standard) and 3 µL internal volume, 0.12 mm i.d. capillary (high-flow). |
|
Valve options | 1 x integrated valve drive as option 2 x external valve drives as option to host user-exchangeable Quick Change valve heads (up to 1300 bar) of different formats: 2-position/6-port, 2-position/10-port, 4-column selection (4-position/10-port) up to 800 bar, 6-column selection (6-position/14-port), 8-column selection (8-position/18-port). Valve heads are automatically identified by their RFID tag. |
|
Column identification | Optionally, column identification kit to track history of up to eight columns. Mounted on left- and right-hand side of module. |
|
Instrument control | LC and CE Drivers A.02.11 or above Instrument Control Framework (ICF) A.02.04 or above Lab Advisor B.02.06 or above InfinityLab Assist (G7180A) with firmware D.07.40 or above Agilent Instant Pilot (G4208A) B.02.19 or above | For details about supported software versions refer to the compatibility matrix of your version of the LC & CE Drivers. |
Communication | Controller Area Network (CAN) | G7116B is a hosted module. (The LC stack needs to contain suitable host module or a LAN card for communication and control). |
Maintenance and safety-related features | Extensive diagnostics, error detection and display (through Instant Pilot control module and Agilent Lab Advisor), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in main maintenance areas. Door-open sensor. |
|
GLP features | Valve heads carrying tags with serial number, pressure rating, number of switches and valve type. Concept of column identification. |
|
Housing | All materials recyclable |
|