Type | Specification | Comments |
---|---|---|
Weight | 12.5 kg (27.6 lbs) |
|
Dimensions (height × width × depth) | 160 x 435 x 436 mm (6.3 x 17.1 x 17.2 inches), Width with column identification kit: 460 mm | G7116A only has a Column ID tag reader on the left hand side. |
Line voltage | 100–240 V~, ±10% | Wide-ranging capability |
Line frequency | 50 or 60 Hz, ±5% |
|
Power consumption | 150 VA, 150 W |
|
Ambient operating temperature | 4–55 °C (39–131 °F) |
|
Ambient non-operating temperature | -40–70 °C (-40–158 °F) |
|
Humidity | < 95% r.h. at 40 °C (104 °F) | Non-condensing |
Operating altitude | Up to 3000 m (9842 ft) |
|
Safety standards: IEC, EN, CSA, UL | Overvoltage category II, Pollution degree 2 | For indoor use only |
ISM Classification | ISM Group 1 Class B | According to CISPR 11 |
Type | Specification | Comments |
---|---|---|
Operating principle | Thermostatted column compartment with dual, independent Peltier-element. Solvent pre-heating and still-air operation for reduction of chromatographic band-broadening under UHPLC-conditions. |
|
Designed for use with Agilent InfinityLab Assist | Intuitive User Interface, Automated Workflows, Predictive Maintenance & Assisted Troubleshooting |
|
Temperature range | 10 °C below ambient (minimum 4 °C) to 85 °C | All specifications are valid for distilled water at ambient temperature |
Temperature range increment | Settable in steps of 0.1 °C | All specifications are valid for distilled water at ambient temperature |
Temperature stability | ± 0.1 °C | All specifications are valid for distilled water at ambient temperature |
Temperature accuracy | ± 0.5 °C (with calibration at 40 °C) | All specifications are valid for distilled water at ambient temperature |
Temperature precision | 0.05 °C | All specifications are valid for distilled water at ambient temperature |
Independent temperature zones | 2 in single device |
|
Column capacity | 4 columns of up to 300 mm length plus InfinityLab Quick Connect fittings or pre-column. The number of pre-column Quick Connect Heat Exchangers is scalable - each column can be equipped with an individual heat exchanger for best performance. 4-column selector valve is available to access each column without replumbing. |
|
Heat-up/cool-down time | 5 min from ambient to 40 °C 10 min from 40 °C to 20 °C <25 min from 25 °C to 85 °C | All specifications are valid for distilled water at ambient temperature |
Solvent heat exchangers | Individually quick-installable for every column. Available at 3 µL internal volume, 0.17 mm i.d. capillary (large ID), 6 µL internal volume, 0.17 mm i.d. capillary (large ID high flow) and 9 µL internal volume, 0.17 mm i.d. capillary (bio-inert, metal-free). |
|
Valve options | 1 x integrated valve drive as option to host user-exchangeable Quick Change valve heads (up to 800 bar) of different formats: 2-position/6-port, 2-position/10-port, 4-column selection (4-position/10-port). Also available in bio-inert materials. Valve heads are automatically identified by their RFID tag. |
|
Column identification | Optionally, column identification kit to track history of up to four columns. Mounted on the left hand-side of module. |
|
Instrument control | LC and CE Drivers A.02.14 or above. Instrument Control Framework (ICF) A.02.04 or above Lab Advisor B.02.08 or above InfinityLab Assist (G7180A) with firmware D.07.40 or above Agilent Instant Pilot (G4208A) B.02.20 or above. | For details about supported software versions refer to the compatibility matrix of your version of the LC & CE Drivers |
Communication | Controller Area Network (CAN) | G7116A is a hosted module. (The LC stack needs to contain suitable host module or a LAN card for communication and control). |
Maintenance and safety-related features | Extensive diagnostics, error detection and display (through Instant Pilot control module and Agilent Lab Advisor), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas. |
|
GLP features | Valve heads carrying tags with serial number, pressure rating, number of switches and valve type. Concept of column identification. |
|
Housing | All materials recyclable. |
|
[1]All specifications are valid for distilled water at ambient temperature (25 °C), set point at 40 °C and a stable flow range from 0.2 - 5 mL/min. Equilibration time: 10 min.
base-id: 3982867979
id: 45036000256572939