For the Bio-inert LC system, Agilent Technologies uses highest-quality materials in the flow path (also referred to as wetted parts), which are widely accepted by life science scientists, as they are known for optimum inertness to biological samples and ensure best compatibility with common samples and solvents over a wide pH range. Explicitly, the complete flow path is free of stainless steel and free of other alloys containing metals such as iron, nickel, cobalt, chromium, molybdenum, or copper, which can interfere with biological samples. The flow downstream of the sample introduction contains no metals whatsoever.
Module | Materials |
---|---|
Agilent 1260 Infinity III Bio-inert Pump (G5654A) | Titanium, gold, platinum-iridium, ceramic, ruby, PTFE, PEEK |
Agilent 1260 Infinity III Bio-inert Multisampler (G5668A) | Upstream of sample introduction:
Downstream of sample introduction:
|
Agilent 1260 Infinity III Bio-inert Manual Injector (G5628A) | PEEK, ceramic |
Agilent 1260 Infinity III Bio-inert Analytical Fraction Collector (G5664B) | PEEK, ceramic, PTFE |
Bio-inert Flow Cells: |
|
G5615-60022 (Standard flow cell bio-inert, 10 mm, 13 µL, 120 bar (12 MPa) for MWD/DAD, includes 0890-1763 – 0.18 x 1500 mm PEEK capillary and 5063-6591 – PEEK fittings) (for Agilent 1260 Infinity III DAD G7115A, and MWD G7165A) | PEEK, ceramic, sapphire, PTFE |
G5615-60005 (Bio-inert flow cell, 8 µL, 20 bar) (for Agilent 1260 Infinity III FLD G7121A/B) | PEEK, fused silica, PTFE |
Bio-inert Heat Exchangers, Valves and Capillaries: |
|
G7116-60041 (Quick Connect Heat Exchanger Bio-inert) (for Agilent 1260 Infinity III Multicolumn Thermostat G7116A) | PEEK (steel-cladded) |
Bio-inert Valve heads (G4235A, G5631A, G5632A, G5639A) | PEEK, ceramic (Al₂O3 based) |
Bio-inert Connection capillaries | Upstream of sample introduction:
Downstream of sample introduction:
|
NOTE
To ensure optimum biocompatibility of your Bio-inert LC system, do not include non-inert standard modules or parts to the flow path. Do not use any parts that are not labeled as Agilent “Bio-inert”. For solvent compatibility of these materials, see General Information About Solvent/Material Compatibility.
base-id: 3358303627
id: 45035999632008587