Warnings and Cautions
WARNING
Ejection of mobile phase
When using sample loops larger than 100 µL, mobile phase may be ejected from the
needle port as the mobile phase in the sample loop
decompresses.
- Please observe appropriate
safety procedures (for example, goggles, safety gloves and protective clothing)
as described in the material handling and safety data sheet supplied by the
solvent vendor, especially when toxic or hazardous solvents are
used.
WARNING
Splashing of solvent
- When using the Needle Port Cleaner, empty the syringe slowly to prevent
solvent from splashing back at you.
- Please observe
appropriate safety procedures (for example, goggles, safety gloves and
protective clothing) as described in the material handling and safety data
sheet supplied by the solvent vendor, especially when toxic or hazardous
solvents are used.
CAUTION
Potential damage to the valve
- Rinse the valve
with water after using buffer solutions to prevent crystals from forming, which
can cause scratches on the rotor seal.
See Flushing the Manual Injector.