The Agilent 1260 Infinity III Bio-inert Manual Injector (G5628A) can be used for manual operation or use of large injection volumes. It offers a standard injection volume of 20 µL (optional 5 µL to 5 mL) and ensures highest injection accuracy.
It uses a Bio-inert 6-port sample injection valve. Sample is loaded into the external 20 µL sample loop through the injection port at the front of the valve. The valve has a PEEK injection seal. A make-before-break passage in the stator ensures that flow is not interrupted when the valve is switched between the INJECT and LOAD positions, and back again.
The valve is mounted on a steel mounting pole, and can be installed at the left- or right-hand side of the LC system.