The ELS Detector has 10 onboard preset methods. These methods comprise evaporator and nebulizer temperatures and gas flow, which can be optimized for specific applications.
These 10 onboard methods are selected using the front keypad and screen, via the METHOD option.
In addition to the 10 on-board preset methods, the ELS Detector has a method XXX that allows modification of the ELSD parameters to be made without the need for software control. Method XXX allows the detector to be used in standalone mode via the front screen and keypad.
To load one of the 10 onboard methods, highlight METHOD. Using the arrow keys scroll up or down to the required method number. The instrument will acknowledge the change by displaying the method number in the top-left corner. These on-board methods cannot be edited.