where
1 | Front screen display |
2 | Keypad |
3 | Eluent inlet |
4 | Solvent waste outlet |
where
1 | Vapor Sensor vent |
2 | Serial RS232 port |
3 | LAN connector (only active on G7102A) |
4 | Firmware button |
5 | I/O connector (Remote Start input) |
6 | Mains Switch |
7 | Mains Input |
8 | Gas Inlet port |
9 | Exhaust port |
| I/O description | Pin number |
---|---|---|
Inputs | Timetable Start | 14 & ground |
| Injection Sync | 13 & ground |
| Remote A/Z | 7 & ground |
Output | Pump stop contact closure – normally open | 3 & 10 |
| Ground (to case) | 1, 5, 6, 11 |
NOTE
To make appropriate remote start and A/Z connections from a third-party LC, a third-party remote start cable for Dimension software (PL0890-0350 (Remote start cable) ) can be purchased from Agilent Technologies.