Standard practice is to add one component at a time back into the HPLC system so that the component causing the problem is easily identified if/when the condition re-occurs.
Begin troubleshooting by adding the pump to the ELS Detector first and finish by adding the column last. If another type of detector is available, use it before the ELS Detector to aid in troubleshooting.