Probable cause | Suggested actions | |
---|---|---|
1 | ERROR: empty node id: 5220443403 | Check the detector settings, such as peak width, and optimize them if needed |
2 | Electrical interference | Ensure that the power supply to the LC system is stable. Check for local interfering sources. If possible, use an independent electrical current for the LC System. |
3 | ERROR: empty node id: 5220453259 | Verify the solvent properties are constant over time. |
4 | Degas mobile phases. Purge the pump and system with a viscous mobile phase to remove any air bubbles. | |
5 | Ensure that the method settings are properly set and the gradient system is delivering the correct gradient composition. Check if the mobile phase reservoirs are empty or if the filter frits are blocked. Check for air bubbles in the system. If necessary, purge the pump and system with a viscous mobile phase to remove any air bubbles. Check the system for any leaks. Check for leaking pump seals. If necessary, replace pump seals. Check for sticking check valves. If necessary, sonicate or replace check valves. | |
6 | Detector cell contamination | Inspect the flow cell for salt crystals, air bubbles, or contamination. If necessary, clean/flush the flow cell or replace it. |