Probable cause | Suggested actions | |
---|---|---|
1 |
Negative drift
Positive drift
Positive drift | |
2 | Room temperature fluctuation | Ensure the system is positioned in a laboratory location free of temperature fluctuations. Use a column thermostat to control column temperature. |
3 | Mobile phase miscibility | Ensure that the mobile phases are completely miscible. |
4 | Replace the column if necessary. | |
5 | Detector cell contamination | Inspect the flow cell for salt crystals, air bubbles, or contamination. If necessary, clean/flush the flow cell or replace it. |
6 | Aging detector lamp | Check if the on-time hours of the detector lamp have exceeded the expected lifetime. Run detector lamp intensity test to check the condition of the lamp. Replace the lamp if necessary. |