When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-2412 | Hex key 2.5 mm, 15 cm long, straight handle |
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]()
|
| Screwdriver |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 5067-6791 | Pressure sensor 1300 bar |
1 | ![]() | 5067-7010 | Pressure sensor biocompatible 1300 bar |
Prerequisites
|
NOTE
This procedure describes how to replace the pressure sensor.
In case the cable to the sensor shall be replaced as well, please contact your Agilent service representative.
Remove capillary connections between the pressure sensor and purge valve.
Remove the screws that fix the pressure sensor to the chassis.
Carefully pull out the pressure sensor for about 2 cm. Then unscrew the cable from the pressure sensor.
Connect the new pressure sensor to the pressure sensor connector.
Fix the pressure sensor to the instrument chassis.
Connect the capillaries from the valve to the pressure sensor: connect port 3 to the pressure sensor inlet and port 2 to the outlet.