Error ID: 22055, 29146
The counter for the solvent volume has exceeded the limit, which has been set in the user interface.
Parameter:
Without Solvent Selection Valve:
0 for channel A, 1 for channel B
With Solvent Selection Valve:
2 for channel A1, 3 for channel B1, 4 for channel A2, 5 for channel B2
Probable cause | Suggested actions | |
---|---|---|
1 | No solvent present. | Refill solvent bottle. |
2 | Inappropriate setting. | Check solvent counter setting in user interface. |
base-id: 4202801163