Error ID: 22031
The target pressure of the binary pump degasser has not been reached within the expected time.
Parameter: Pressure in mbar
Probable cause | Suggested actions | |
---|---|---|
1 | Condensation in degasser chamber due to temperature fluctuation. | Equilibrate and restart module. Use Evacuation Mode if necessary. |
2 | Degasser is defect. | Please contact your Agilent service representative. |
base-id: 4202796683