Error ID: 22434
The current of the secondary pump drive is too high.
Parameter: 2 or 3 referring to pump drive
Probable cause | Suggested actions | |
---|---|---|
1 | Blockage of flow path between secondary pump head and pressure sensor, e.g. of the heat exchanger. | Check for blockages in the flow path. Please contact your Agilent service representative. |
2 | Secondary pump drive is defect. | Please contact your Agilent service representative. |
base-id: 4202694027
id: 4202694027