Error ID: 22433
The current of the primary pump drive is too high.
Parameter: 1 or 4 referring to pump drive.
Probable cause | Suggested actions | |
---|---|---|
1 | Blockage of flow path between primary pump head and pressure sensor, e.g. of the heat exchanger. | Check for blockages in the flow path. Please contact your Agilent service representative. |
2 | Primary pump drive is defect. | Please contact your Agilent service representative. |