Error ID: 22407, 29205
The current consumption of the pump drive is too low.
Parameter: 1 – 4 referring to pump drive
Probable cause | Suggested actions | |
---|---|---|
1 | Drive motor defect. | Please contact your Agilent service representative. |
2 | Wrong/missing connection of pump drive to main board. | Please contact your Agilent service representative. |
base-id: 4202525195
id: 4202525195