Error ID: 22406, 29214
Pump drive blocked or encoder failed.
Parameter: None
Probable cause | Suggested actions | |
---|---|---|
1 | Blockage of system before pressure sensor. | Check for blockage of e.g. outlet valve filter frit, purge valve, heat exchanger. |
2 | Drive encoder failed. | Please contact your Agilent service representative. |
base-id: 4202499595