Error ID: 22064, 29164
The flow rate of the binary pump has exceeded the limit, while the pump runs in pressure controlled mode, e.g. during a pressure test.
Parameter: None
Probable cause | Suggested actions | |
---|---|---|
1 | Leak | Check for leaks in the pump and flow path. |
2 | Bottle empty | Check bottle filling. Fill solvent bottle. |
3 | Shutoff valve closed. | Open shutoff valve. |
4 | Drift of pressure sensor (unlikely for short tests taking only minutes). | Replace pressure sensor. |